Pharmaceutical packaging for biologics: extractables, leachables and container-closure integrity
What FDA's August 2026 draft container closure guidance, the ICH Q3E draft, FDA's integrity-testing guidance and Annex 1 say about leachables, materials, integrity testing and packaging change control for biologics, as of 27 November 2026.
For a small-molecule tablet, packaging is mostly a barrier. For a protein in solution it is part of the formulation. The container sits against the product for its whole shelf life, and a change in glass, polymer, stopper, lubricant or sterilisation step can change what the protein sees. This article draws on the FDA draft guidance Container Closure Systems for Human Drugs and Biological Products (August 2026), the ICH Q3E draft guideline on extractables and leachables (Step 2, consultation 2025), the FDA's 2008 guidance on container and closure system integrity testing, the EU GMP Annex 1, and a 2017 PDA Letter article on prefilled syringe compatibility.
Last reviewed: 27 November 2026. The FDA container closure guidance and ICH Q3E are drafts, and the PDA article is an industry article, not guidance. Passages marked as interpretation are the author's analysis.
Extractables and leachables: the definitions and the risk
ICH Q3E's draft describes an extractable study as a process by which chemical entities are extracted from a test article, with solvents and conditions chosen to reflect the worst case of manufacturing or storage. Extractables that could become leachables, meaning substances that actually migrate into the drug product, are then assessed. The draft sets out a quality risk management process, following ICH Q9, with a risk assessment, risk control and lifecycle management started for every product. It frames risk as multidimensional: on the quality side, the interaction between component and formulation, the properties and pretreatment of the component, storage and manufacturing conditions (including surface area to volume ratio, temperature and contact time), and the leaching propensity of the formulation, including pH, co-solvents and surfactants. On the safety side it lists route of administration, patient population, maximum dose, dosing frequency and treatment duration.
For biological products, the draft adds that risk identification and mitigation should include evaluating interactions between reactive leachables and formulation components that may affect quality, safety or efficacy, with possible modification mechanisms such as denaturation, aggregation or degradation considered. It also notes that although leaching is minimal in the frozen state, the storage component should be evaluated before freezing and after thawing. Liquids are described as higher risk than non-lyophilised solids because of their greater leaching propensity.
On patient risk, the draft sorts leachables into classes. Class 1 covers compounds that need control below the usual qualification threshold, including the ICH M7 cohort of concern and compounds with potent non-mutagenic toxicity such as bisphenol A or benzo(a)pyrene; the draft says the practical approach is to avoid materials that may leach them. Class 3 covers compounds with low systemic toxicity potential, and Class 2 sits between. Organic leachables above the analytical evaluation threshold (AET) should be identified, quantified and reported for safety assessment.
Materials: what each contact surface can bring
FDA's August 2026 draft lists compromised packaging performance as a risk factor, giving glass delamination as an example, and says that processes such as washing, coating, lyophilisation and sterilisation of the container closure system can affect it, with delamination possibly resulting in particulate matter. It also points to multiple-dose products, where the elastomeric closure is pierced repeatedly, and refers to USP General Chapter <381> on elastomeric components. It notes that an elastomeric stopper may interact with an organic solvent in a formulation to raise leachables, and that adsorption of the drug by the container closure system can affect the product.
The 2017 PDA Letter article on prefilled syringes states that Type I glass remains the primary material for syringe barrels, with cyclic olefin polymer (COP) and cyclic olefin copolymer (COC) plastic syringes gaining popularity, and that multiple elastomer formulations exist for plungers and tip caps. It identifies silicone oil and tungsten residues as compatibility issues and reports published cases in which soluble tungsten residue was the root cause of aggregation in clinical-trial batches of epoetin alfa and in an alpha-helical protein formulation at about pH 4.0, while a monoclonal antibody study showed rapid precipitation by tungsten at pH 5.0 and lower risk at pH above 6.0. Its recommended screening design spikes the formulation with sodium tungstate at 0, 1, 10 and 100 ppm and compares samples by size-exclusion HPLC and particle analysis, and for silicone oil it recommends flow imaging because droplet size can overlap with protein particles and the droplets have a characteristic spherical appearance.
Interpretation: the materials comparison has no single winner. Glass, COP and elastomers each shift the risk, from delamination and particles to oxygen and moisture behaviour to extractables and lubricant. Which surface is best depends on the formulation's pH, buffer, surfactant and protein, so the compatibility data should drive the choice, not the other way round.
Study design: from prior knowledge to simulation
ICH Q3E's draft starts with prior knowledge: composition, food-contact compliance, statements that particular compounds were not added, compendial tests, existing extractables studies, biological reactivity testing, pretreatment (sterilisation, cleaning, flushing, siliconisation, surface treatments) and prior use history. Extractables studies should use a product-specific AET, test the component as finally processed, use solvents of varied pH and polarity relevant to the formulation (including excipients and surfactants), cover volatile, semi-volatile, non-volatile and elemental extractables, and use multiple complementary analytical techniques. A semi-quantitative study can precede a leachables study, with an analytical uncertainty factor applied in the AET calculation. A leachables-to-extractables correlation should be established in addition to the risk assessment. The supplier's extractables report may be supplemented with additional testing, and the manufacturer is responsible for setting requirements for its unique product.
Container closure integrity
FDA's 2008 guidance accepts container and closure system integrity testing in place of sterility testing as part of a stability protocol, and lists properly validated physical or chemical tests (bubble tests, pressure or vacuum decay, trace gas permeation or leak tests, dye penetration, seal force, electrical conductivity and capacitance) and microbiological tests (microbial challenge or immersion). It names limits of sterility testing: it detects only viable organisms present at the time of test, can give false positives from testing contamination, and is destructive. Under EU Annex 1, containers closed by fusion, such as blow-fill-seal, should be tested at 100% with validated methods for small volumes (up to 100 ml), with reduced sampling allowed for larger ones; vacuum-sealed containers should be tested for vacuum maintenance before release and during shelf life; and integrity validation should consider transport and shipping effects such as decompression or extreme temperatures.
Interpretation: the regulators' lists do not rank methods. A practical reading is that the method should be validated for the specific container, the defect size that matters and the product, used across the shelf life, and chosen with its sensitivity and the nature of the product in mind (for example, whether a liquid biologic can tolerate the test conditions).
Vials, prefilled syringes and other presentations
FDA's question-based review for sterility assurance of terminally sterilised products asks sponsors to list every container and closure component of each drug product configuration, with description and manufacturer, and to include every proposed commercial manufacturer. Its own example is that if vials are bought from two manufacturers, both are listed. The vial and stopper example is a table of vial, stopper and seal. A prefilled syringe is listed by barrel, plunger or stopper, plunger rod, cap and needle shield, with parts that do not touch the product identified. For complex systems the description should cover the fluid pathway, meaning the components in direct contact with the dosage form.
The same document asks that integrity validation address every interface that works as a sterile barrier and the whole fluid pathway. Acceptable methods it lists include microbial immersion challenge, dye ingress and egress, yttrium ingress, helium leak testing, pressure decay and high voltage leak detection. Ideally every container and closure system proposed for commercial production is represented in the study. FDA gives a vial example: if two vials with identical neck dimensions are used with the same stopper, one vial neck and stopper interface needs to be included. The study units should also have seen the production conditions that could affect integrity, such as maximum sterilisation or depyrogenation and multiple exposures.
The sources read for this article do not describe cartridges as a separate presentation. A cartridge is a glass barrel with a plunger and a closure at the other end, so the syringe questions on silicone oil, tungsten and plunger compatibility are the nearest evidence, but that is an inference, not a statement from the sources.
Supplier qualification and the lifecycle of a packaging change
ICH Q3E's draft lists changes that should trigger a fresh look: new information on a material, formulation changes (for example surfactant composition or concentration, which can change both the type and amount of leachables), changes in composition, supplier, process, geometry or pretreatment of contact materials, changes to manufacturing process conditions such as solvent system, temperature, pH, cleaning or sterilisation, flushing, surface area to volume ratio, and changes that affect patient exposure such as dose, duration, route or population. FDA's draft guidance also has a section on reporting postapproval changes to a container closure system.
Interpretation: for a biologic, the useful change-control question is not only whether the new component meets its specification but whether the protein sees a different surface, lubricant or leachable profile. A supplier change that looks neutral on paper can still alter siliconisation or an elastomer's extractables, which is why Q3E lists supplier and pretreatment changes by name.
Illustrative scenario
This is a made-up example, not a named company. A monoclonal antibody in a prefilled syringe moves to a plunger from a second supplier. The team compares the supplier's extractables data, repeats a leachables screen with the product formulation including its surfactant, runs a silicone-oil and agitation study with subvisible particle and size-exclusion analysis, and repeats integrity testing at the end of shelf life before filing the change.
Questions to ask before a packaging change (interpretation)
- Which dimension of risk changes: material, formulation, process, dose or population?
- Does the extractables study use the real formulation, including surfactant and pH, and worst-case contact?
- Has the protein been tested against silicone oil, tungsten or other relevant residues?
- Is the integrity method validated for this container, and is it applied through shelf life and shipping?
- Does the change need a regulatory variation or supplement?